Troubleshooting, Troubleshooting 89 – MTS Criterion 60 - Manual User Manual

Page 89

Advertising
background image

MTS Criterion™ Series 60 Product Manual

Troubleshooting

89

Troubleshooting

Basic Troubleshooting

S

YMPTOM

C

HECK

R

EASON

Cannot clear interlocks,
no apparent source for
the interlock.

Is any Emergency Stop button

pressed?

Any pressed Emergency Stop will

prevent interlocks from being

cleared.

You can now check if an ESTOP is

pressed in the fault status screen in
TestWorks.

Is the pump on?

On a MTS Criterion Series 60 the

pump must be on to clear interlocks.

Is there a fault on either the Interface

indicator on the controller box or
MTS Criterion Series 60 logic

indicator on the IOP panel? Check
the indicator LEDs or check the

software fault status.

Any fault that is detected will prevent

machine operation. Generally contact
MTS tech support or field service to

troubleshoot. Faults are reset when
power is cycled.

Is something plugged into the guard
connector in the J11 connector?

Make sure the guard connection is

not plugged into the adjacent aux
power connector.

The guard connector contains
jumpers to configure the guard as

either “slow” or “interlock”. When

nothing is plugged in, the guard gets
configured as interlock and the guard

appears open.

Is something plugged into the J3
connector on the MTS Insight

controller board?

Although the interlock connection on
the MTS Insight board is rarely used,

if a jumper plug is not installed an

interlock will be generated.

Is the guard door open and guard

configured to Interlock?

The guard function “slow” or

“interlock” is configured by jumpers

in the cable. If no jumpers are
installed, it is configured to interlock

when the guard door is open.