Operation

Travel Limit Switches (Physical Limits)

There are three limit switches on MTS Criterion Series 60 frames, the crosshead
upper limit, crosshead lower limit, and piston travel limit.

Crosshead travel limits

They are normally closed and are activated when the limit switch contacts the
limit plate.

Piston travel limit

The piston travel limit is normally closed and is activated when the piston
reaches the end of stroke and the limit switch contacts the limit plate.

Piston Travel Limit

Crosshead
Lower Limit

Crosshead
Upper Limit

Adjusting the position
of this sleeve to
change the Crosshead
Lower Limnit

Adjusting the position of
this block to change the
Crosshead Upper Limit