Additional Digital I/O Information

The digital inputs have an MOCD223 optical isolator with 2.7 Kohm, ½ Watt
series resistor. To reliably turn on they need 1 mA of current. This means the
minimum input high voltage is 4.0 V DC. The maximum input voltage is 28.0 V
DC. The device should be off for input voltages less than 1.0 V.

The digital outputs are implemented by an AQV252G PhotoMOS relay with a
0.75 Amp poly fuse in series. Although the device is rated at 60 V peak, but it is
recommended that a maximum of 48 V be applied. If the load is highly inductive,
such as a relay coil, an appropriate snubber network should be used near the coil
terminals to prevent large flyback voltages from exceeding the device ratings.

A typical example might be connecting an external switch.