Controller Connections
MTS Criterion™ Series 60 Product Manual
Installation
65
J2 Handset
This is intended to interface to the handset. Specifics for this connector are:
•
12 V output power with 200 mA current limit
•
RS422 driver (differential)
•
RS422 receiver (differential)
•
Interlock input. Handset shorts between INTLK+ and INTLK- when it is
connected.
•
8-pin RJ-45 connector
J3 Interlock
This is intended to connect to a test area enclosure switch that opens when the
door is opened. If not used, a jumper plug (p/n 049-635-901) must be installed. If
you are building a cable, the maximum length is 30.48 m (100 ft) with 24 gauge
wire. The switch should be wired normally closed, such that when the switch
opens, an interlock is generated.
8-Pin Assignment
P
IN
S
IGNAL
1
Transmit +
2
Transmit -
3
+12V
4
INTLK +
5
INTLK -
6
Analog GND
7
Receive +
8
Receive -