J14 emergency stop – MTS Criterion 60 - Manual User Manual
Page 70
MTS Criterion™ Series 60 Product Manual
70
Controller Connections
Installation
J14 Emergency Stop
This connector is intended to connect to a remote Emergency Stop switch. If not
used, a jumper plug must be installed. If you are building a cable, the maximum
length is 30.48 m (100 ft) with 18 AWG. The switch should be wired normally
closed, such that when the switch is pressed, an interlock is generated.
Pin assignment is as follows:
ESTOP3 = emergency stop switch with two sets of contacts (optional remotely
mounted).
ESTOP_OUT = optional ESTOP output for accessory, includes two contact
outputs and one feedback.
7
No Contact
8
VCC
9
DIN 1-
10
DIN 2-
11
DIN 3-
12
OUT 1-
13
OUT 2-
14
OUT 3-
15
AGND
Pin Assignment (Continued)
P
IN
S
IGNAL
Pin Assignment
P
IN
S
IGNAL
1
ESTOPB_OUT-
2
No Contact
3
ESTOP_OUT_MONITOR-
4
ESTOP_OUT_MONITOR+
5
ESTOP3B+
6
ESTOPA_OUT-
7
ESTOP3B-
8
ESTOP3A+
9
ESTOPB_OUT+
10
No Contact
11
No Contact
12
No Contact
13
ESTOP3A-