MTS Criterion™ Series 60 Product Manual

70

Controller Connections

Installation

J14 Emergency Stop

This connector is intended to connect to a remote Emergency Stop switch. If not
used, a jumper plug must be installed. If you are building a cable, the maximum
length is 30.48 m (100 ft) with 18 AWG. The switch should be wired normally
closed, such that when the switch is pressed, an interlock is generated.
Pin assignment is as follows:
ESTOP3 = emergency stop switch with two sets of contacts (optional remotely
mounted).
ESTOP_OUT = optional ESTOP output for accessory, includes two contact

outputs and one feedback.

7

No Contact

8

VCC

9

DIN 1-

10

DIN 2-

11

DIN 3-

12

OUT 1-

13

OUT 2-

14

OUT 3-

15

AGND

Pin Assignment (Continued)

P

IN

S

IGNAL

Pin Assignment

P

IN

S

IGNAL

1

ESTOPB_OUT-

2

No Contact

3

ESTOP_OUT_MONITOR-

4

ESTOP_OUT_MONITOR+

5

ESTOP3B+

6

ESTOPA_OUT-

7

ESTOP3B-

8

ESTOP3A+

9

ESTOPB_OUT+

10

No Contact

11

No Contact

12

No Contact

13

ESTOP3A-