Chapter 3 Display Operations

Menu Operations

Series PM172 Powermeters

47

6. Press ENTER to confirm your new setting, or press ESC

to discard changes.

To exit the menu, press ESC.

Counters Setup

The PM172 has four six-digit counters that can count pulses delivered

through the device digital inputs with a programmable scale factor, or events
that trigger setpoint operations. This menu allows you to link digital inputs to
the counters and define a pulse multiplier for each counter. To enter the
menu, select the “Cnt” entry from the Main menu and press the ENTER

button.

The menu uses three entries:
1. The upper window indicates a counter number.
2. The middle window selects a digital input to be linked to

the counter.

3. The lower window defines a counter multiplier.

Use the UP and DOWN arrow buttons to scroll to the desired counter.

To change the counter options:

1. Press the SELECT button to activate the middle window.
2. Use the UP and DOWN arrow buttons to link a digital

input to the counter or to disable the counter input.

3. Press the SELECT button to activate the lower window.
4. Use the UP and DOWN arrow buttons to enter the desired

multiplier.

5. Press ENTER to confirm your changes and to store your

new settings, or press ESC to discard changes.

6. You are returned to the upper window to select another

counter or exit the menu.

To exit the menu, press ESC.

The following table lists available counter options.

Parameter Options Default

Description

Pulse source

None = disabled
Inp.1 = DI1
Inp.2 = DI2

None

Links a digital input to the counter

Multiplier

1-9999

1

The value added to the counter
when a pulse is detected on the
pulse source input, or the counter is
incremented through a setpoint
action

Relay Output Setup

To enter the menu, select the “rEL” entry from the Main menu and press the

ENTER button.

Use the UP and DOWN arrow buttons to scroll to the desired relay.

To select a relay parameter:

1. Press the SELECT button to activate the middle window.
2. Use the UP and DOWN arrow buttons to scroll to the

desired parameter.

To change the parameter value:

1. Press the SELECT button to activate the lower window.

Cnt.1

1

Inp.1

rEL

ESC