Number of Records per Series = Sampling Rate (Samples per Cycle) x
Number of Cycles per Event / 512

The total number of records you must allocate to store the

required number of events (series) is defined as follows:

Number of Records = Number of Records per Series x Number of Series

For example, if you want to record a 64-cycle waveform
sampled at a rate of 32 samples per cycle, the number of
records required for one waveform series would be:

Number of Records per Series = (32 x 64)/512 = 4.

If you want to allocate space sufficient to store 20
waveform events (series), you should set up the
waveform log file for 4 x 20 = 80 records.

4. Click OK, and send your setup to the meter or save to the

database.

5. Click “Setup Recorder”, or click on the “Waveform

Recorder” tab.