3-48 IETG FlowHawk Manual

Securing the other end of the band

Note: If the sensor mounting plate has too much side
play, attach another anchor bolt approximately 25 mm
from the plate.

10. Slide the sensor mounting plate along the metal band until it is

level (from side to side) with the flow surface and pipe crown,
and tighten the nuts securely with a nut driver. Refer to the
instructions on leveling this sensor in Installing the Ring in the
Pipe
beginning on 3-25.

11. Install the sensor onto the mounting plate in the following way

based on the sensor type:

Surface Combo Sensor Mount this sensor by

inserting the studs on the back of the sensor into the
keyholes on the mounting plate and sliding the sensor
back (in a downstream direction) until it is fully in place.
The sensor cable should exit the downstream edge of the
band.