Maintenance and troubleshooting 8-37 – ADS Environmental Services IETG FlowHawk QR 775012 A4 User Manual

Page 321

Advertising
background image

Maintenance and Troubleshooting 8-37

1

L

E

N

N

A

H

C

L

E

N

N

A

H

C

M

AX

Y

8

L

2

N

k

O

P

T

S

E

T

a

.

W W W. IETG

U K

.CO .

CO M M

2

FLOWHAWK

TM

IETG

Ex ia IIB T4 Ga

Sira 09ATEX2027X

II 1 G

0518

MODEL: 8000 - FHK/FST

Use only battery pack 106152

Returning the monitor top to the proper orientation with the canister

14. Establish communication with the monitor remotely or on site

using the IS PC Communication Cable (p/n 8000-0054) to
verify communication. Refer to Connecting Directly to the
Monitor
in Chapter 4, Communication, for instructions on
attaching the IS PC Communication cable to the monitor and
laptop computer.

15. Finger-tighten the top bolts to make sure the top and gasket are

seated properly. Then, tighten the bolts about halfway to full
torque value by tightening the opposite bolts. Finally, tighten
the bolts to approximately 3.39 Nm (30 in-lbs/2.50 ft-lbs) to
fully secure the top to the canister. If the monitor will not seal,
loosen the bolts and check for a trapped wire or debris on either
the gasket or the lip. You can increase the torque value on the
bolts incrementally to assist in sealing. However, tighten the
bolts evenly to prevent distorting the gasket and do not exceed
5.0843Nm (45-inch pounds)
.

Advertising