ADS Environmental Services IETG FlowHawk QR 775012 A4 User Manual

Page 335

Advertising
background image

Maintenance and Troubleshooting 8-51

Problem

Individual ultrasonic depth readings (SDEPTH_3) from
the Surface Combo Sensor are consistently equivalent
to the value obtained from the following equation: Pipe
Height – Physical Offset + 9.99.

Possible

Causes

Sensor may be faulty.

5-volt fuse on regulator may have blown.

Possible

Solutions

Replace the Surface Combo Sensor.

Replace the 5-volt blue fuse on the regulator or return
the monitor to IETG for regulator replacement.

Caution: Do not connect a potentially shorted sensor to
another monitor until confirming that the sensor is
functional (i.e., not shorted) using a voltmeter. A
shorted sensor will blow the 5-volt (blue) replacement
fuse on the regulator of the monitor.

Problem

Sensor reports abnormal temperatures.

Possible

Causes

Sensor connection to monitor may be loose.

Temperature sensor may be faulty.

Sensor may be faulty.

Possible

Solutions

Secure sensor connection to monitor if necessary.

Replace the sensor.