ADS Environmental Services IETG FlowHawk QR 775012 A4 User Manual

Page 332

Advertising
background image

8-48 IETG FlowHawk Manual

Problem

Range from ultrasonic depth subsystem is significantly
greater than manually measured range.

Possible

Causes

Physical offset may be incorrect.

Pipe height may be incorrect.

Electronic offset may be incorrect.

Sensor may not be level.

Foam or other substance may be absorbing the pulse.

Sensor may be faulty.

Possible

Solutions

Verify the physical offset.

Verify the pipe height.

Verify electronic offset and adjust if necessary.

Make sure the sensor is level and in good condition.

Check the hydraulic conditions in the pipe.

Fire the sensor at a shorter distance onto a hard surface
to confirm accuracy.

Replace the sensor.

Contact your regional IETG representative.