Configuration and Activation 6-51

Electronic Offset Enter the difference that exists between
the smart depth sensor readings and the manually measured
depth, when applicable. The default value is 0.

Note: An electronic offset is rarely needed. An offset
greater than +/- 13 millimeters typically indicates that the
physical offset or pipe height has been measured or
recorded incorrectly.

Mode Select the mode in which you want the smart depth
sensor to operate based on environmental factors or site
constraints. Although the current default setting is High
Power
, IETG recommends using the Normal setting under
most conditions. However, if erratic Smart Depth readings
occur, consider using the High Power setting.

Store Data Select this checkbox to ensure the monitor logs
the smart depth data to memory. This option is selected by
default.

Advanced Select this button to access the Advanced
Parameters
dialog.

Advanced Parameters dialog

Note: Modifications made to the Advanced Parameters
should be performed only by or under the guidance of
trained IETG personnel. Therefore, IETG recommends
leaving the default settings (Auto) in place.