Maintenance and Troubleshooting 8-53

Problem

UpDepth data seems erratic over a full depth range.

Possible

Causes

Sensor transmit/receive surfaces may be dirty.

Sensor may be rotated greater than 15 degrees from
center.

Sensor may be faulty.

Processor board may be faulty.

Possible

Solutions

Clean the sensor transmit/receive surfaces.

Check the sensor connections.

Ensure the sensor is installed as closely as possible to
the 6:00 position in the pipe, silt allowing.

Contact your regional IETG representative.

Replace the sensor.

Replace the monitor if defective.

Problem

UpDepth readings are abnormally low.

Possible

Causes

Sensor transmit/receive surfaces may be covered with
silt.

Sensor may be physically offset in the flow and the
correct physical offset has not been measured and/or
downloaded to the monitor.

Possible

Solutions

Re-measure physical offset of the sensor.

Check the site conditions, and relocate the sensor if
necessary.