1-6 IETG FlowHawk Manual

Note: The 8K-CS1 is no longer available; however, IETG
will continue to provide support for this sensor.

• As aluminum is used at the accessible surface of this unit, in the

event of rare incidents, ignition sources due to impact and
friction sparks may occur. This shall be considered when it is
being installed, particularly in locations that require equipment
with a Group II Ga level of protection.

• The only communication devices that may be connected to the

COMM connector on the FlowHawk are the GSM Wireless
Module (p/n 8000-0052), the EMU/EMUX Communication
Cable (p/n 106227B), and the Direct Connect Interface (p/n
8000-0054)

Note: The Triton-Telog Comm Cable (p/n 8000-0054-01)
also may be connected to the COMM connector on the
FlowHawk to provide communication between the Telog
Ru-33 and the monitor. However, since the Ru-33 is not
certified for IS operation, an installation involving a
connection between these units will not be considered
intrinsically safe, unless the Telog unit is installed outside
the hazardous area.

Never rub any plastic parts (specifically, the ABS plastic

connector potting boots and the Peak Combo Sensor and
Slimline Peak Combo Sensor housings) with a dry cloth. This
could result in the buildup of dangerous static charges.

• Parts of the enclosure may be non-conducting and may generate

an ignition-capable level of electrostatic charge under certain
extreme conditions. Therefore, do not install this equipment in
a location where it may be subject to external conditions (such
as high-pressure steam or dust) that may cause a buildup of
electrostatic charge on non-conducting surfaces.