3 detailed operation, Initial start-up, Self-diagnosis – Bio-Rad Model 680 Microplate Reader Accessories User Manual

Page 22: Initialization, Security check, System ready

Advertising
background image

18

4.3 Detailed

Operation

4.3.1. Initial Start-up






































01:Kinetic-1

M405(1)R605(6)30read

Shake:999s,Mid,Every

38.2/40.2 18/09/01

End point mode

Kinetic mode

Self-diagnosis

Model 680
Microplate Reader

13:05 18/09/01

System Login

User: Administrator

Password:

*****

Press ENTER

Login SCREEN

01:EndPoint Assay01

M405(1)R655(6)
Shake:999s,Mid

38.2/40.2 18/09/01

Checkmark

Validation Program

Incu:OFF 18/09/01

Checkmark mode

Initializing reader

MAIN SCREEN

Self-Diagnosis

This screen will appear upon power-up and during self-diagnosis. Date and Time shall
also be indicated on this screen.

Initialization

After self-diagnosis, the system shall initialize all the hardware.

Security Check

The system requires input of Password to confirm that the user is the authorized
operator.
Pressing the Change/Right arrow key will switch the user mode between Common
user and Administrator.
Password for both Common user and Administrator is set as “00000” initially upon
factory shipment.
Operator can change the password in the EDIT menu. See Security later.

System Ready

When this MAIN screen is on the LCD, Start/Stop key is activated for microplate
reading. Pressing Start/Stop key will begin reading according to the setting of the
current active protocol.
There are three kind of MAIN screen depending on the active protocol type: Endpoint,
Kinetic, and Checkmark program.
Four arrow keys, Numeral keys, Main key and Enter key are inactive.
Start/Stop, Memory Recall, Edit, Print and Paper Feed keys are active.

01:EndPoint Assay01

M405(1)R655(6)

Shake:999s,Mid

38.2/40.2 18/09/01

Kit name for current active protocol

01:Kinetic-1

M405(1)R605(6)30read

Shake:999s,Mid,Every

38.2/40.2 18/09/01

Kinetic

M: Measurement wavelength and its position in the filter disk.
R: Reference wavelength and its position in the filter disk.
Number of reading for KINETIC

Temperature setting (if available)

Temperature reading if Incubator activated

Storage number in memory of
current active protocol

Plate shaking parameters: period in seconds and speed

M: Measurement wavelength and its position in the filter disk.
R: Reference wavelength and its position in the filter disk.

Date (day / month / year)

Plate shaking parameters:

period in seconds
speed (Fast, Middle, Slow)
mode (before every read, before first read and no shake)

End point