Bio-Rad Model 680 Microplate Reader Accessories User Manual

Page 65

Advertising
background image

61

Abs. Data Sample of <Mes. data> and <Ref. data>
" 0.101 0.102 0.103 0.104 0.105 0.106 0.107 0.108 0.109 0.110 0.111 0.112"<",">
" 0.201 0.202 0.203 0.204 0.205 0.206 0.207 0.208 0.209 0.210 0.211 0.212"<",">
" 0.301 0.302 0.303 0.304 0.305 0.306 0.307 0.308 0.309 0.310 0.311 0.312"<",">
" 0.401 0.402 0.403 0.404 0.405 0.406 0.407 0.408 0.409 0.410 0.411 0.412"<",">
" 0.501 0.502 0.503 0.504 0.505 0.506 0.507 0.508 0.509 0.510 0.511 0.512"<",">
" 0.601 0.602 0.603 0.604 0.605 0.606 0.607 0.608 0.609 0.610 0.611 0.612"<",">
" 0.701 0.702 0.703 0.704 0.705 0.706 0.707 0.708 0.709 0.710 0.711 0.712"<",">
" 0.801 0.802 0.803 0.804 0.805 0.806 0.807 0.808 0.809 0.810 0.811 0.812"

Example:
eia.reader DRPD ,0,1,End point #1 ,0,415, ,2, ,1,03/2/18 14:00:00,begin, 0.101 0.102 0.103 0.104 0.105
0.106 0.107 0.108 0.109 0.110 0.111 0.112, 0.201 0.202 0.203 0.204 0.205 0.206 0.207 0.208 0.209
0.210 0.211 0.212, 0.301 0.302 0.303 0.304 0.305 0.306 0.307 0.308 0.309 0.310 0.311 0.312, 0.401
0.402 0.403 0.404 0.405 0.406 0.407 0.408 0.409 0.410 0.411 0.412, 0.501 0.502 0.503 0.504 0.505
0.506 0.507 0.508 0.509 0.510 0.511 0.512, 0.601 0.602 0.603 0.604 0.605 0.606 0.607 0.608 0.609
0.610 0.611 0.612, 0.701 0.702 0.703 0.704 0.705 0.706 0.707 0.708 0.709 0.710 0.711 0.712, 0.801
0.802 0.803 0.804 0.805 0.806 0.807 0.808 0.809 0.810 0.811 0.812,end,


1-7-2. Kinetic's Download Raw Plate Data Format

The assignment of the Download Raw Plate Data format of Kinetic:

#

Item of the plate datum

Details

1 Plate data mode

“0”(End point), “1”(Kinetic)

2 Memory number

"1" to "10" (End point), "1" or "2" (Kinetic)

3 Total number of reads

"2" to "30"

4 Reading number

"0", "1" to "30"
(1). The number 0 means that the successive data from here are

the reading condition.

(2). The number 1 to 30 means the reading number of the series of

Kinetic readings.

It should not exceed the total number of reads.

(1) Reading number = "0"

5 Kit name [16]

ASCII characters (max 15characters)

6 Plate reading mode (S/D)

Plate reading mode (“0”: Single / “1”: Dual)

7 Measurement wavelength

Model 680: “400” to “750”nm

8 Reference wavelength

Model 680: “400” to “750”nm
A space character (0x20) is available for Single reading.

9 Measurement filter number

“1” to “8”

10 Reference filter number

“1” to “8”
A space character (0x20) is available for Single reading.

11 Protocol number

"1" to "64" (End point), "65" or "66" (Kinetic)

12 Shaking mode

"0"(disable), "1"(before 1st read), "2"(before every read)

13 Shaking speed

"0"(low), "1"(Middle), "2"(High)

14 Shaking time

"0" to "999" sec

15 Start delay

"0" to "999" sec

16 Reading interval [29]

"6" to "1029" sec for Model 680
Interval time values must be separated by a space code.
1-2,2-3,3-4,4-5,5-6,6-7,7-8,8-9,9-10,10-11,11-12,12-13,13-14,14-1
5,15-16,16-17,17-18,18-19,19-20,20-21,21-22,22-23,23-24,24-25,
25-26,26-27,27-28,28-29,29-30