Bio-Rad Model 680 Microplate Reader Accessories User Manual
Page 53

49
1-2-6 RM (Reset maintenance report)
<command>: "RM"
Reset maintenance report values.
<arguments>: none
<response>: "ERE"<space><error
code><cr>
1-2-7. ISTATUS (Get incubator status)
<command>: "ISTATUS"
Request incubator status from reader.
<arguments>: none
<response>: "ERE"<space><error
code><space><inc
status><cr>
<inc status>:
<incubator exist status><space><on/off><space><setting temperature value>
<space><sensor temperature value>
incubator exist status: “0”:The reader don’t have incubator / “1”:The reader have
incubator.
on/off - Incubator ON/OFF status, "0": incubator OFF / "1": incubator ON
setting & sensor temperature value:
(Temperature is 3 bytes ASCII encoded decimal value and floating point. ex: "37.0")
1-2-8. INCU (Set Incubator)
<command>: “INCU“
Incubator ON/OFF control; Benchmark Compatible Format
<arguments>: <on/off>
<cmd args>:
on/off - Incubator ON/OFF
"0"; incubator OFF
"1"; incubator ON
<response>: "ERE"<space><error
code><cr>
1-2-9. WINC (Set Incubator Temperature)
<command>: “WINC”
Set Incubator control temperature data
<arguments>:
< temperature >
<cmd args>:
Setting temperature value (degree).
ASCII encoded decimal(1-10)
temperature set range: 25.0 to 45.0 degree
Incubation adjustable in increments of 0.1 degree.
<response>: "ERE"<space><error
code><cr>
1-2-10. FSTATUS (Get filter wheel status)
<command>: "FSTATUS"
Request filter wheel status from the reader.
<arguments>: none
<response>: "ERE"<space><error
code><cr><filter
status><cr>
<filter status>:
<wl.1><space><wl.2><space><wl.3><space><wl.4><space>
<wl.5><space><wl.6><space><wl.7><space><wl.8>
wl.1 – wl.8:
filter wavelength at position 1 – position 8. ex.”405”(ASCII encoded decimal)
“***” will be returned when filter is not installed