1-5.

“TSE” and “TSE eq” Assay Parameter Format


“TSE” and “TSE eq” parameter is same format.
All parameter data is ASCII code data.
Transmit format: <“,”(0x2c)><TSE item 1><“,”><TSE item 2><“,”> …<TSE item 7><“,”>
“,”(0x2c) delimits each parameters item.

“TSE” and “TSE eq” assay data download or upload orders:

Item

P.C or Reader Transmit data

1

Report title [24]:

Report title (for internal & external print report)

ASCII characters (max 24characters)

2

Maximum OD value for Control
Negative

“1” to “3500”

This value is Maximum OD * 1000, no decimal point
data length max: 4 characters

3

Minimum O.D value for Control
Negative

“-3500” to “3500”

This value is Minimum OD * 1000, no decimal point

data length max: 5 characters (include sign character)
(This value is lower than Max O.D value for CN.)
This value is not being used.

4

validation data rang for Control
Negative

“1” to “999”

Data range (0.001 to 0.999 * 1000, no decimal point
data length max: 3 characters

5

Minimum O.D value for Control Positive

“1” to “3500”

This value is Minimum OD * 1000, no decimal point
data length max: 4 characters include sign flag

6

Calculation value k1
(Positive Cut-off=(CNx + K1) * k2)

“0” to “9999”

K1 Calculation value * 1000
data length max: 4 characters

7

Calculation value k2
(Positive Cut-off=(CNx + K1) * k2)

“0” to “9999”

K2 Calculation value * 1000
data length max: 4 characters

8

Calculation value k3
(Negative Cut-off= (CNx + K3) * k4)

“0” to “9999”

K3 Calculation value * 1000
data length max: 4 characters

9

Calculation value k4
(Negative Cut-off= (CNx + K3) * k4)

“0” to “9999”

K4 Calculation value * 1000

data length max: 4 characters


Example:
“,TSE assay report ,2500,123,400,1010,210,1000,210,900