Bio-Rad Model 680 Microplate Reader Accessories User Manual

Page 63

Advertising
background image

59

1-6-2-2. Full Data Type Format for Kinetic

<upload raw plate data>: <memory #><data exist><date time><plate #>< <reading info> | <plate info>

<Mes. data><cr>
[<Ref. data><cr>] > --- Ref.data present only for dual wavelength reading.

<memory #><data exist><date time>: Same as "1-6-1. Header Type Format of the Upload Raw Plate Data"

<plate #>: 0 to 30

"0": It only generates <reading info>.
"1" to "30": this number generates each plate data.


(1). <plate #>= 0
<reading info>: "Protocol #: "<protocol number><cr>

<protocol number>: 65 to 66

"Kit name: "<kit name><cr>

<kit name>: Max 15 char

"Shaking mode: "<shaking mode><cr>

<shaking mode>: "No" / "Before 1sr read" / "Before every read"

"Shaking speed: "<shaking speed><cr>

<shaking speed>: "Low" / "Middle" / "High"

"Shaking time: "<shaking time><cr>

<shaking time>: 0 to 999

"Number of reads: "<# of reads><cr>

<# of reads>: 2 to 30

"Read interval: "<interval 1><space><interval 2><space>… <interval 29><cr>

interval time: 6 to 1029

“Mes. filter: ”<measurement filter wavelength data><space>"("<filter number>")"<cr>

<wavelength data>: 400 to 750 for Model 680

<filter number>: 1 to 8

“Ref. filter:”<reference filter wavelength data><space>"("<filter number>")"<cr>

Only when Dual wavelength reading


(2). <plate #>= 1 to 30
<plate info>: "Plate #: "<plate number><cr>
<date

time><cr>

<Mes. data><cr>[<Ref. data><cr>]: Same as Abs. Data (See "1-3. Absorbance Data Format")

Advertising