Accuracy

±1.0% or 0.010 from 0 to 3.0 OD at 490nm

Linearity

±1.0% from 0 to 2.0 OD, ±2.0% from 0 to 3.0 OD

Reproducibility

1.0% or 0.005 OD from 0 to 2.0 OD, 1.5% from 2.0 to 3.0 OD

Inter-channel variation

1.5% or 0.005 OD from 0 to 3.0 OD

Stability and Drift

0.010 OD at OD=1 with single wavelength reading

Temperature set
point

25ºC to 45ºC, 0.1ºC step
Set point must be 5ºC above ambient.

Set point accuracy Less than

±0.5ºC when set point =37ºC at RT=25ºC

Well to well
uniformity

Less than 0.7ºC when set point =37ºC at RT=25ºC

Incubator
(Factory option)

Warm-up time

Less than 5 minutes when set point =37ºC at RT=25ºC

Internal Printer
(Factory option, external
printer port is not
available when this
option is installed)

Thermal printer, Paper Width 112mm

Line Voltage

Universal input: 100 - 240 VAC, 50/60Hz

Power consumption

100 VA max.

Dimensions

340(W) x 330(D) x 152(H) mm

Weight 5.5

kg

Operating Environment 5ºC to 35ºC

0 to 95% relative humidity without condensation

Storage Temperature

-20ºC to 50ºC
0 to 95% relative humidity without condensation