c. Recall STD curve setting

Recall Setting of Standard Curve

In this screen, there are two items for setting the Recall STD curve as below.
The Up and Down arrow key moves the cursor to the desired item in the menu
list.

1. Recall setting (Yes/No)
2. Selecting the storage number of standard curves (1 -> 5)


Recall setting

Move the cursor to the desired position using the Up, Down
or Right arrow keys. Enter key accepts the setting and
returns to the Standard menu screen.

Yes: Recalling of the standard curve is active.
No: Recalling of the standard curve is inactive.


Storage number selection

Five memories are selectable. The vacant memory which is
marked with asterisks on the date and time field is also
selectable. But it is necessary to save any standard curve
before using its protocol.
Move the cursor to the desired position using the Up and
Down
keys.


This setting screen consists of two parts. The Right arrow key
is used to switch the indication field between the date and
time to the standard curve name.
Move the cursor to the desired position using the Up and
Down keys. Enter key accepts the storage number and
returns to the Standard menu screen.




NOTE:

If the recall of standard curve is active (the recall setting is
“Yes”), the Standard information setting and the Curve fit
setting will be invalidated and there setting screens will not
open.


Recall STD curve

»■Recall setting

•Select STD curve

Recall setting

»■Yes No

Recall setting

Select STD Curve

»■1:16/03/04 11:25

2:******** *****

↓ 3:15/03/04 10:30

Select STD Curve

»■4:******** *****

5:******** *****

Select STD Curve

»■1: STD Curve #1

2: STD Curve #2

↓ 3: STD Curve #3

Select STD Curve

»■4: STD Curve #4

5: STD Curve #5

Select STD curve

Right arrow key