Bio-Rad Model 680 Microplate Reader Accessories User Manual

Page 58

Advertising
background image

54

1-4. Protocol

data

All protocol data is ASCII code data.
Transmit format:

<“,”(0x2c)><protocol data 1>< “,”><protocol data 2><“,”>…<protocol

data 42><”,”>

”,”(0x2c) delimits each protocol item.

Protocol data download or upload orders:

Item

P.C or Reader Transmit data

1

Protocol mode

“0”(endpoint) or “1”(kinetic), “2”(TeSeE), “3”(TeSeE eq)

2

Security lock status

“0” unlock / “1” lock

3

Kit name[16]:

ASCII characters (max 15characters + null code)

4

Plate reading mode

Plate reading mode (“0”:Single / “1”:Dual)

5

Evaluation mode(endpoint only)

“1”(Normal) or “2”(Evaluation)

6

Precision mode

1:Standard precision mode / 2:High precision mode

7

Number of reads

“1” (endpoint) / “2” to “30”(kinetic mode)

8

Interval time [30](Kinetic mode)

[0]:Start delay time, [1] to [29]:interval time
Start delay: “0” to “999” sec
Interval time: “0” to “1029” sec

9

Measurement wavelength data

“400” to “750”nm

10

Reference wavelength data

“400” to “750”nm
(When the single reading mode is single, Reference wavelength
should be set to “ “(0x20).)

11

Shaking mode

“0” to “2”
“0”(disable)
“1”(1

st

read(kinetic mode only))

“2”(before every read)

12

Shaking speed

“0” to “2”
“0”(Low) / “1”(Middle) / “2”(High)

13

Shaking time

“0”to”999”sec

14

Incubation control temperature

“25.0” to“45.0” (include decimal point)

15

Incubator enabled

“0” or “1”
“0”(No) / “1”(Yes)

16

Cutoff type

“0” to “4”
“0”(Don’t use cutoff)
“1”(cutoff constant)
“2”(cutoff control)
“3”(value by formula)
“4”(Ratio to calibrator)

17

Cutoff value

“0” to “1”
“0”(Single cutoff)
“1”(Ranged cutoff)

18

Cutoff formula type

“1” to “12”
“1”: k*CNx “7”: k + CNx
“2”: k*CPx “8”: k + CPx
“3”: k*COx “9”: k +COx
“4”: CNx / k “10”: k*CNx + CPx
“5”: CPx / k “11”: (CNx+CPx) / k
“6”: COx / k “12”: k1*CNx + k2*CPx

19

Cutoff unit<cr>

“0”(STD unit) / “1”(OD)

20

Cutoff K (K and K1)

“0.01” to “10.0”
(data length max: 4 characters(include decimal point))

21

Cutoff K2

“0.01” to “10.0”
(data length max: 4 characters(include decimal point))

22

Cutoff positive const value

“.001 to 9999”

Advertising