Section 7. Specifications

7.1 Instrument

Specifications

Operating panel

Membrane pad with 4 arrow keys, 7 function keys and 11 numeral keys

Display

4 lines x 20 alphabet characters LCD

Computer interface

Bi-directional RS232C serial communication port, 9 pin D-sub plug

External printer port
(Only when the internal
printer option is not
installed )

Centronics interface with ASCII character code
25 pin D-sub receptacle
80 characters output per line

Report types

Raw, Absorbance, Matrix, Limit, Cutoff, Concentration, curve fit, difference,
Kinetic Report

Applicable plates

96 well microtiter plate of Polystyrene “flat”, ”U”, ”V” bottom rigid type
(Plate Footprint: 127 x 85mm(standard size))
8 and 12 well strip plate
Maximum plate height: 16 mm

Mixing

3 speeds( Low, Mid, High) Programmable time period: 0 - 999 seconds

Data storage

Latest 10 data plates for end-point, latest 2 data plates of 30 readings for
kinetic readings, protocol data and calendar/clock function shall be kept in
memory by a lithium battery even when the power is off.
Average lifetime of the battery shall be 5 years.

Warm-up time

3 minutes

Photometric
Methods

Single and dual wavelength

Photo detectors

8 silicon photodiodes for measurement and 1 silicon photodiode for reference

Light source

Tungsten halogen lamp 20W with 3000 hours average lifetime

Spectral range

400nm to 750nm

Interference filters

Wavelength tolerance:

±3.0nm, Band width: 10nm typically

Filter capacity

8 filters

Standard filters

450, 415, 490, 655nm

Optional filters

405, 540, 550, 570, 595, 630, 750nm

Reading speed

Fast Reading

6 second with single wavelength reading
10 second with dual wavelength reading

Standard Reading

15 second with single wavelength reading
30 second with dual wavelength reading

Indication range

0.000 to 3.5 OD

Resolution 0.001

OD