d. Cutoff Ratio setting

Ratio to Calibrator

In this screen, there are three items for setting the ratio to calibrator. The Up and
Down
arrow key moves the cursor to the desired item in the menu list. The Enter
key seals the selection and resumes to the next setting screen.


Concentration of Calibrator

The Numerical keys and the Decimal point key are used to
enter the calibrator concentration value. The Enter key sets
the value and returns to the previous screen.


Unit of Calibrator

17 units are supported, as for the standard's unit. The unit
marked with “[ ]” is the current selection.
The setting procedure is the same as the unit selection of
Standards.


Cutoff value setting

This menu screen selects Single or Ranged cutoff type. The
cutoff type marked with “[ ]” is the current selection. The
Enter key seals the selection and resumes to the next setting
screen


Single Cutoff Control

The Numerical keys are used to enter the Positive cutoff
concentration value and gray zone value. The Enter key sets
these cutoff ratio values and returns to the previous screen.


Ranged Cutoff Control

The Numerical keys are used to input the Positive and the
Negative cutoff concentration value. The Enter key sets these
cutoff ratio and returns to the previous screen.


NOTE:

The calibrator concentration value is used to convert the
absorbance value of each well into concentration value. The
ratio of Concentration/Absorbance of Calibrator well is
applied to the conversion. The cutoff is done with the defined
Positive and Negative values to the converted concentration
value of each well.


Ratio to Calibrator
»■Concentration

•Unit

•Cutoff value

Concentration:

9999

unit: mol/l

»■ 01: ug/dl
02: ng/ml

03:[ pg/ml]

↓ 04: pm/ml

»■Positive >=

9999

Gray zone:

50

%

Unit: mol/l

»■Positive >=

9999

Negative <

9999

Negative<Positive

Unit: mol/l

»■Ranged cutoff

[Single cutoff]

Concentration

Unit

Cutoff value

Single cutoff

Ranged cutoff