3 command language for model 680 microplate reader – Bio-Rad Model 680 Microplate Reader Accessories User Manual

Page 52

Advertising
background image

48

7.3

Command Language for Model 680 Microplate Reader

7.3.1 Command Language for Remote control mode


1-1. Syntax

SYNTAX:
Command:

<device name><space><command>[ <space><command args.>... ]<cr>

<device name>:

"EIA.READER" – Model680 reader

Response:

<response header><space><error code>[ <space><response data>... ]<cr>

<response header>: “ERE”

1-2. Commands
1-2-1. ID (Get Instrument ID)

<command>: "ID"

Request device ID

<arguments>: None
<response>: "ERE"<space><error

code><space><id><cr>

<id>: “Model

680


1-2-2. AQ (Go to remote control mode)

<command>: "AQ"

Acquire remote control of the device and lock front panel key pad.

<arguments>: none
<response>: "ERE"<space><error

code><cr>

NOTE: Until the reader receives the “AQ” command, the reader will not accept
any other command by responding error code of “8073”-Device not in remote mode.
Once the reader received the “AQ” command, the reader will turn into remote mote
and lock all keys on front panel except for START/STOP for release.


1-2-3. RL (Release remote control mode)

<command>: "RL"

Release remote control of the device and lock front panel key pad.

<arguments>: no

arguments

<response>: "ERE"<space><error

code><cr>


1-2-4. RS (Reset device)

<command>: "RS"

Reset device to power up configuration local mode

<arguments>: none
<response>: "ERE"<space><error

code><cr>

1-2-5. MR (Maintenance report)

<command>: "MR"

Transmit maintenance report

<arguments>: none
<response>: "ERE"<space><error

code><cr><mrecords><cr>

<mrecords>: maintenance

records;

“On/Off:” <#on/off><cr>:number of times the reader was turned on.
“Hours :” <#hours><cr>:number of hours the reader was on.
“Plates:” <#plates><cr>:number of plates read.
Note: <#on/off>,<#hours> and <#plates> are 4 ASCII encoded decimal values.
This information is kept in battery backup RAM

Advertising