1-7.

Download Raw Plate Data Format

All protocol data is ASCII code data.
Transmit format: <“,”(0x2c)><plate datum 1><“,”><plate datum 2><“,”>…<plate datum n><“,”>
“,”(0x2c) delimits each parameters item.

1-7-1. End point's Download Raw Plate Data Format

The assignment of the Download Raw Plate Data format of End point:

#

Item of the plate datum

Details

1 Plate data mode

“0”(End point), “1”(Kinetic)

2 Memory number

"1" to "10" (End point), "1" or "2" (Kinetic)

3 Kit name [16]

ASCII characters (max 15characters + null code)

4 Plate reading mode (S/D)

Plate reading mode (“0”: Single / “1”: Dual)

5 Measurement wavelength

Model 680: “400” to “750”nm

6 Reference wavelength

Model 680: “400” to “750”nm
A space character (0x20) is available for Single reading.

7 Measurement filter number

“1” to “8”

8 Reference filter number

“1” to “8”
A space character (0x20) is available for Single reading.

9 Protocol number

"1" to "64" (End point), "65" or "66" (Kinetic)

10 Plate reading date

<year>”/”<month>“/”<day><space><hour>”:”<minutes>“:”<seconds>
<year>: "00" to "99" (2000 to 2099)
<month>: "1" to "12"
<day>: "1" to "31"
<hour>: "0" to "23"
<minutes>: "0" to "59"
<seconds>: "0" to "59"

11 Begin of <Mes. data>

"begin"

12 <Mes. data>

<Mes.data>: Abs. Data
(See below "Abs. Data Sample")
Measurement wavelength absorbance values are in ASCII encoded
floating point format. Ex. 0.000
Each absorbance value is separated by a space, but in case of minus
value, a minus sign is put instead. Each row of 12 well values is
separated by a comma (,).
<8 rows of 12 wells of abs. values>:
Row1=wells (A1 to A12)"," Row2=wells (B1 to B12)","…
…. Row8=wells (H1 to H12)

13 End of <Mes. data>

"end"

14 Begin of <Ref. data>

"begin", put only for dual wavelength reading

15 <Ref. data>

<Ref.data>: Abs. data, put only for dual wavelength reading
(See below "Abs. Data Sample")
<8 rows of 12 wells of abs. values>:
Row1=wells (A1 to A12)"," Row2=wells (B1 to B12)","…
…. Row8=wells (H1 to H12)

16 End of <Ref. data>