63

7.3.2 Error

Codes

Error

code

Error Cause

Recovery

0000

No error

8071

Invalid Command

The received data is not a command for the
instrument

8072

Parameter out of range

The received parameter is out of range

8073

Device Not In Remote
Mode

The instrument is not in remote mode

8074

Device busy

The instrument is doing a job.

8075

Filter wheel

Filter wheel jammed.

“reset” command.

8076

8077

Light bulb burned out

Insufficient light level

“reset” command.

8078

8079

8080

8081

8082

8083

Incubator error

Incubator over heat.

8084

Plate carrier error-1

Plate carrier home sensor closed status error

“reset” command.

8085

Plate carrier error-2

Plate carrier home sensor open status error

“reset” command.

8086

Plate carrier error-3

Plate carrier jammed before reading

“reset” command.

8087

Plate carrier Position

Plate carrier jammed after reading

“reset” command.

8088

Light level error

There is too big difference of light between
channels

“reset” command.

8089

Filter missing

Light is exceeding allowable range

“reset” command.

8090

Disagreement of security
status

The security status from P.C was disagreement.

8091

RS232C error

RS232C hardware error

8092

A/D converter error

Malfunction of A/D converter.

“reset” command.

8093

Measurement stop by stop
key

The user input “stop” key during measurement.

“reset” command.







Advertising