Bio-Rad Model 680 Microplate Reader Accessories User Manual

Page 42

Advertising
background image

38

Warning and Error of Automatic Mapping:

Warning screen:
If the total number of mapping wells exceeds 96 wells, a warning message in the
left figure will appear on the screen. The calculation of the total number of
mapping wells is as below:

(Blanks+Standards+CNs+CWs+CPs+COs+QCs+Calibrators+Sample)*Replicates


The plate mapping is executed normally and the exceeded part is invalidated.

Error screen:
If one or more mapping parameters are over the input range, an error message in
the left figure will appear on the screen.
On the third line of the LCD, the name of the parameter which has been out of
range and its input range will appear.

Example: “BLK: 0 – 10”

In case of more than one error, the first parameter of errors is indicated. By
pressing nay key it returns to the automatic mapping screen.

Example of automatic mapping:

An example of automatic mapping is as below.
1. Conditions of each parameter

Blank = 1
Standard = 8
Control CN = 2, CW = 0, CP = 2, CO = 0, QC = 0
Calibrator = 1
Sample = 32
Replicate number = 2


2. Plate mapping data

1 2 3 4 5 6 7 8 9 10 11 12
A
B
C
D
E
F
G
H

B
B
S1
S1
S2
S2
S3
S3

S4
S4
S5
S5
S6
S6
S7
S7

S8
S8
CN1
CN1
CN2
CN2
CP1
CP1

CP2
CP2
CL1
CL1
X1
X1
X2
X2

X3
X3
X4
X4
X5
X5
X6
X6

X7
X7
X8
X8
X9
X9
X10
X10

X11
X11
X12
X12
X13
X13
X14
X14

X15
X15
X16
X16
X17
X17
X18
X18

X19
X19
X20
X20
X21
X21
X22
X22

X23
X23
X24
X24
X25
X25
X26
X26

X27
X27
X28
X28
X29
X29
X30
X30

X31
X31
X32
X32
E
E
E
E



Warning: The total

Of mapped wells has
Exceeded 96.

Press any key

Entered value is

Out of range

BLK: 0 – 10

Press any key

Advertising