1-2-11. RWELL (Read a plate well)

<command>: “RWELL

Read a plate well

<arguments>: Single

reading:<cl><space><rw><space><wp1>

Dual reading: <cl><space><rw><space><wp1><space><wp2>

< cmd args>:

<cl>: Column number ASCII encoded decimal(1-12)
<rw>: Row number ASCII encoded decimal(1-8)
<wp1>: Measurement filter position ASCII encoded decimal(1 to 8)
<wp2>: Reference filter position ASCII encoded decimal(1 to 8)
Optional; only present for Dual wavelength reading

<response>: "ERE"<space><error

code><space><abs.val><space>

<optional abs.val2 for reference wavelength><cr>

<abs.val1>: Absorbance values at measurement wavelength

(ASCII encoded floating point format) ex. “1.234”

<abs.val2>: Absorbance

values at reference wavelength

(ASCII encoded floating point format) ex. “1.234”


1-2-12. RPLATE (Read plate)

<command>: "RPLATE"

Perform plate reading once

<arguments>: <sp><speed><sp><mix time><sp><mix speed><sp><meas. filter><sp><ref. filter>
< cmd args>:

<speed>: Plate reading speed, “1”: fast speed, “2”: Normal speed( ASCII encoded
decimal)

<mix time>:

Plate mixing time(0 to 999) (ASCII encoded decimal)

<mix speed>:

Plate mixing power(“L”:Low, “M”:middle,”H”:High)

<meas. filter>:

Measurement filter position(1 to 8)(ASCII encoded decimal)

<ref. filter>:

Reference filter position(1 to 8)(ASCII encoded decimal)
(Optional; only present for Dual wavelength reading)

<sp>: space(‘

‘)

<response>: ”ERE”<space><error

code><space><abs.

data><cr>

1-2-13. RTPLATE (Retransmit plate data)

<command>: "RTPLATE"

Retransmit the last one plate data
If the previous reading was read in the kinetic mode, the last plate data means the
final plate data from the set of plate data.

<arguments>: none
<response>: ”ERE”<space><error

code><space><abs.

data><cr>