Bio-Rad Model 680 Microplate Reader Accessories User Manual
Page 62

58
1-6.
Upload Raw Data Format
1-6-1. Header Type Format of the Upload Raw Plate Data
<upload raw plate data>: <memory #><data exist><date time><cr>
< memory #>: (1). End point
“End point # " <memory number>”<cr>
<memory number>: 1 to 10 (End point)
ex. “End point #1”
(2). Kinetic
“Kinetic # " <memory number>”<cr>
1 or 2 (Kinetic)
<data exist>: "Exist: "<exist><cr>
<exist>: Exist = "1", No exist= "0"
ex. “Exist: 1”
<date time>: <day> ”/” <month> “/” <year> <space> <hour> ”:” <minutes> “:” <seconds> ><cr>
ex. “01/01/2003 01:05:02”
1-6-2. Full Data Type Format of the Upload Raw Plate Data
1-6-2-1. Full Data Type Format for End point
<upload raw plate data>: <memory #><data exist><date time><reading info>
<Mes. data><cr>
[<Ref. data><cr>] --- Ref.data present only for dual wavelength reading.
<memory #><data exist><date time>: Same as "1-6-1. Header Type Format of the Upload Raw Plate Data"
<reading info>: "Protocol #: "<protocol number><cr>
<protocol number>: 1 to 64
"Kit name: "<kit name><cr>
<kit name>: Max 15 char
“Mes. filter:”<measurement filter wavelength data><space>"("<filter number>")"<cr>
<wavelength data>: 400 to 750
<filter number>: 1 to 8
“Ref. filter:”<reference filter wavelength data><space>"("<filter number>")"<cr>
Only when Dual wavelength reading
<Mes. data><cr>[<Ref. data><cr>]: Same as Abs. Data (See "1-3. Absorbance Data Format")