caUtion: Do not cover filter perforated section with paper wrapped
instruments or any other objects that could obstruct the way to air flow and
steam flow. Permanent damage may occur to the container as a result of this
obstruction.

ProcessinG - HiGH vacUUm (Pre-vacUUm)
Run loaded sterilizer according to the manufacturer’s instructions at:

• High Vacuum (pre-vacuum, three pulse, standard)
• Temp: 270° F
• Exposure Time: 4 Minutes (minimum)
• Cycle Dry Time: 20 Minutes (minimum)
• Cool Time: Varies according to load contents

caUtion: Cool drafts from air ducts or other air currents should be avoided
during the cooling phase to avoid post-sterilization moisture caused by rapid
cooling syndrome
.

• The container needs to remain on the container cart, in a draft free area
until cool enough to handle.

note: A Container with a solid bottom may require additional cooling time;
the additional required cooling time to be determined by the user.

sHelf life
A real-time event-related shelf life study had been performed on the
containers. The conclusion of the study is that the container is capable of
storing high vacuum (pre-vacuum) sterilized medical devices for a maximum
of 360 days for the Solid Bottom Containers and 360 days for the
Perforated Bottom Containers as long as the container integrity has not
been compromised.

tHe miltex container system Warranty
Miltex hereby warrants all Miltex Containers purchased directly from Miltex or
an Authorized Dealer in accordance with the following terms and conditions:

The Miltex Container is guaranteed to be free from defects in material and
workmanship when used for its intended purpose. Any Miltex Container
that proves defective in workmanship or material will either be repaired or
replaced, at Miltex discretion, without charge.

Containers not purchased from an authorized source may not be authentic or
meet Miltex quality standards.

container Gaskets Warranty
The warranty period for Miltex container gaskets is three years from the date
of purchase.

Periodic maintenance
Periodic Maintenance is suggested to ensure all parts are in good
working order.

repair or replacement
Under this Warranty, and at its discretion, Miltex will repair or replace any
parts of the product found to be defective in workmanship or material.
Miltex will determine, at its discretion, the cause and nature of any defect,
the necessity and manner of repair or replacement, and all other matters
pertaining to the conditions of the Miltex Container System.
Miltex, Inc. Returned Goods Policy applies to all returned items (refer to
details in current Surgical price list).

exclusions
This Warranty shall not apply to any conditions(s) or damage resulting from
negligence, misuse, improper handling, improper cleaning, improper opening
techniques, or unauthorized repair work including but not limited to:
• Incidents of abuse such as denting of the Miltex Container due to dropping
or other instances of mechanically applied pressure.
• Damage determined to be related to caustic or abrasive cleaning agents.
• Items modified by the customer.
• Items modified or customized by Miltex at the request of the customer.
• Damage from fire, flood and other occurrences not under the control
of Miltex, Inc.

Please direct your questions concerning this Warranty to:
Miltex, Inc., 589 Davies Dr., York, PA 17402
www.miltex.com

0310/0/C/4

SURG-CONT-INST

phone 717 840-9335