KEYENCE LJ-V7000 Series User Manual

Page 12

Advertising
background image

12

LJ-V7000_COM_RM_E

Name

Setting item designation structure

Definition

Typedef struct {

BYTE

byType;

BYTE

byCategory;

BYTE

byItem;

BYTE

reserve;

BYTE

byTarget1;

BYTE

byTarget2;

BYTE

byTarget3;

BYTE

byTarget4;

} LJV7IF_TARGET_SETTING;

Description

Information for specifying target setting items.
byType, byCategory, byItem

When modifying or reading a setting, these variables are used to specify the target setting
item.

byTarget1, byTarget2, byTarget3, byTarget4

These variables are used when specifying further details for the setting item.
For example, when configuring OUT measurement mode, these are used to specify the
OUT number.

Comment

For details, see the appendix.

Name

Measurement results structure

Definition

Typedef struct {

BYTE

byDataInfo;

BYTE

byJudge;

BYTE

reserve[2];

FLOAT

fValue;

} LJV7IF_MEASURE_DATA;

Description

Measurement value and judgment results.
byDatainfo

This variable indicates whether or not the measurement value (fValue) is valid, and if it is
not a valid value, what kind of data it is. See LJV7IF_MEASURE_DATA_INFO.

byJudge

Tolerance judgment result. See LJV7IF_JUDGE_RESULT.

fValue

Measurement value. The unit used for measurement values is the minimum display unit
set for Target OUT in program settings.
When the minimum display unit is 1 mm to 0.001 mm, the measurement value unit is [mm].
When 1 um to 0.1 um, the measurement value unit is [um]. The unit for sectional areas is
mm

2

, and the unit for angles is deg.

When not a valid value, a large negative value is stored (-10

10

).

Comment

-

Advertising