KEYENCE LJ-V7000 Series User Manual

Page 84

Advertising
background image

84

LJ-V7000_COM_RM_E

2:Knee

byte

Setting Data

0

Area Left: any value in measurement range

(0.01μm unit Sined 32-bit integer

example: 5mm=500000)

1
2
3
4

Area Right: any value in measurement range

(0.01μm unit Sined 32-bit integer

example: 5mm=500000)

5
6
7
8

Knee shape: 0:Valley, 1:Peak

9

Detection direction: 0: + direction, 1: - direction

10

Detection No: 1~10

11

Sensitivity: 0~100

3:Intersection (lines)

byte

Setting Data

0

Linear calculation area: 0: Area2

disabled, 1: Area2 enabled

Line1

1

Reserved (fixed as 0)

2

Reserved (fixed as 0)

3

Reserved (fixed as 0)

4

Line calculation area Left: Any

value in measurement range

(0.01μm unit Sined 32-bit integer.

example: 5mm=500000)

5
6
7
8

Line calculation area Right: Any

value in measurement range

(0.01μm unit Sined 32-bit integer.

example: 5mm=500000)

9

10
11
12

Area2 Left: Any value in

measurement range (0.01μm unit

Sined 32-bit integer.

example: 5mm=500000)

13
14
15
16

Area2 Right: Any value in

measurement range (0.01μm unit

Sined 32-bit integer.

example: 5mm=500000)

17
18

19

20

Linear calculation area count: 0:

Area2 disabled, 1: Area2 enabled

Line2

21

Reserved (fixed as 0)

22

Reserved (fixed as 0)

23

Reserved (fixed as 0)

24

Line calculation area Left: Any

value in measurement range

(0.01μm unit Sined 32-bit integer.

example: 5mm=500000)

25
26
27
28

Line calculation area Right: Any

value in measurement range

(0.01μm unit Sined 32-bit integer.

example: 5mm=500000)

29
30
31
32

Area2 Left: Any value in

measurement range (0.01μm unit

Sined 32-bit integer.

example: 5mm=500000)

33
34
35
36

Area2 Right: Any value in

measurement range (0.01μm unit

Sined 32-bit integer.

example: 5mm=500000)

37
38

39

4:Contact (line-arc)

byte

Setting Data

Linear calculation area: 0: Area2

disabled, 1: Area2 enabled

Line

1

Reserved (fixed as 0)

2

Reserved (fixed as 0)

3

Reserved (fixed as 0)

4

Line calculation area Left: Any

value in measurement range

(0.01μm unit Sined 32-bit integer.

example: 5mm=500000)

5

6

7

8

Line calculation area Right: Any

value in measurement range

(0.01μm unit Sined 32-bit integer.

example: 5mm=500000)

9

10

11

12

Area2 Left: Any value in

measurement range (0.01μm unit

Sined 32-bit integer.

example: 5mm=500000)

13

14

15

16

Area2 Right: Any value in

measurement range (0.01μm unit

Sined 32-bit integer.

example: 5mm=500000)

17

18

19

20

Arc calculation area: 0: Area2

disabled, 1: Area2 enabled

Arc

21

Reserved (fixed as 0)

22

Reserved (fixed as 0)

23

Reserved (fixed as 0)

24

Arc calculation area Left: Any

value in measurement range

(0.01μm unit Sined 32-bit integer.

example: 5mm=500000)

25

26

27

28

Arc calculation area Right: Any

value in measurement range

(0.01μm unit Sined 32-bit integer.

example: 5mm=500000)

29

30

31

32

Area2 Left: Any value in

measurement range (0.01μm unit

Sined 32-bit integer.

example: 5mm=500000)

33

34

35

36

Area2 Right: Any value in

measurement range (0.01μm unit

Sined 32-bit integer.

example: 5mm=500000)

37

38

39

Advertising