9.2.9.4 Detailed setting errors

The settings have a consistency that must be observed. The detailed setting errors (dwError) that
are returned for the send setting and reflect write settings area request commands for settings that
not satisfy this consistency are listed below.

*1: X indicates the program number and 0 to F is stored there.
*2: YY indicates the OUT number and 00 to 0F is stored there.

9.2.9.5 Profile data values

Profile data is output as signed 32-bit data, but for points where the profile could not be correctly
detected, the following values are output.

dwError value

Error details

0x01000000

The operation mode is "high-speed (profile only)" and the memory
allocation is set to entire area (do not overwrite)

0x1X000000 (*1)

The sampling period is outside the configurable range

0x1X06YY00 (*2)

Invalid measurement mode (when set to simple 3D measurements in
batch measurement OUT, etc.)

0x1X06YY01 (*2)

Measurement mode "operator" is set for OUT1

0x1X06YY02 (*2)

When the measurement mode is operator, the operator target OUT is
invalid

0x1X06YY03 (*2)

Measurement value filter cutoff frequency error

0x1X06YY04 (*2)

Prohibited combination in the measuring range settings (when
configured to specify the measuring area when batch is off, etc.)

0x1X06YY05 (*2)

Measuring range setting (when the batch count and the measuring
range do not match (also including simple 3D measurement Y
direction measurement range))

0x1X06YY06 (*2)

When the measuring range setting is reference OUT, the consistency
conditions were not fulfilled

0x1X06YY07 (*2)

Could not match the tolerance setting (upper/lower limit and
hysteresis)

0x1X06YY08 (*2)

Could not match the scaling setting (measurement value 1/2, span
after calculation)

0x1X070000 (*2)

Could not match the analog output scaling setting (OUT display value
1/2, span after calculation)

0x1X080000 (*2)

The amount of storage has exceeded the configurable range

Value (hexadecimal

notation)

Name

Reason

-2147483648
(0x80000000)

Invalid data

This value is output when the peak could not
be detected.

-2147483647
(0x80000001)

This value is output when the data is invalid
due to a setting such as a mask having been
set.

-2147483646
(0x80000002)

Dead zone data

This value is output when the data is located
in a dead zone.
This value is only output when dead zone
processing is enabled.

-2147483645
(0x80000003)

Judgment wait data

This value is output when there is an
insufficient amount of profiles for averaging.