KEYENCE LJ-V7000 Series User Manual

Page 18

Advertising
background image

18

LJ-V7000_COM_RM_E

Name

Advanced mode get profile response structure (batch measurement: on)

Definition

Typedef struct {

DWORD

dwGetBatchNo;

DWORD

dwGetBatchProfCnt;

DWORD

dwGetBatchTopProfNo;

BYTE

byGetProfCnt;

BYTE reserve[3];

} LJV7IF_GET_BATCH_PROFILE_ADVANCE_RSP;

Description

The profile information returned for the get profiles command when the operation mode is
"advanced mode (with OUT measurement)" and batch measurements are on.
dwGetBatchNo

The batch number that was read this time.

dwGetBatchProfCnt

The number of profiles in the batch that was read this time.

dwGetBatchTopProfNo

Indicates what number profile in the batch is the oldest profile out of the
profiles that were read this time.

byReadProfCnt

The number of profiles that were read this time.

Comment

-

Name

Get storage status request structure

Definition

Typedef struct {

DWORD

dwReadArea;

}LJV7IF_GET_ STRAGE_STATUS_REQ;

Description

Get target designation information in the get storage status command.
dwReadArea

The storage surface to read.
• When the memory allocation setting is "double buffer"

0: Active surface, 1: Surface A, 2: Surface B

• When the memory allocation setting is "entire area (overwrite)", fixed as 1
• When the memory allocation setting is "entire area (do not overwrite)"

0: Active surface, surface specification (1 to 999)

Comment

"Active surface" refers to the surface of the buffer that profile data is being written onto. For
further details, refer to "9.2.9.2 Internal memory".

Name

Get storage status response structure

Definition

Typedef struct {

DWORD

dwSurfaceCnt;

DWORD

dwActiveSurface;

} LJV7IF_GET_STRAGE_STATUS_RSP;

Description

The storage status information returned for the get storage status command.
dwSurfaceCnt

Storage surface count

dwActiveSurface

The active storage surface.
When the active program has storage off, 0.

Comment

For details about "Storage surface", refer to "9.2.9.2.2 For operation mode: advanced (with
OUT measurement)".