Parameters, see "measurement area details – KEYENCE LJ-V7000 Series User Manual

Page 83

Advertising
background image

83

LJ-V7000_COM_RM_E

• Storage settings

• Program name

• Measurement area details

The correction target selection of feature point
correction of position correction and the unique
parameters of the measurement target of the
measurement mode are shown below.The byte
numbers shown here indicate the byte numbers from
the start of the corresponding data block. (see the
example at 12.4)

<Storage target>

Type:10h~1Fh (10h:Program NO.0, 11h:Program NO.1,

, 1F:Program NO.15)

Category:08h, Item:01h
Target1~4:00h

byte

Setting Data

0

Storage target: 0:OFF, 1:OUT value, 2:Profile

1

Reserved (fixed as 0)

2

Reserved (fixed as 0)

3

Reserved (fixed as 0)

<Storage condition>

Type:10h~1Fh (10h:Program NO.0, 11h:Program NO.1,

, 1F:Program NO.15)

Category:08h, Item:02h
Target1~4:00h

byte

Setting Data

0

Storage condition: 0:Terminal/Command, 1: OUT

update, 2:OUT data (edge), 3:OUT data (level)

1

Reserved (fixed as 0)

2

Reserved (fixed as 0)

3

Reserved (fixed as 0)

<Storage data amount (Terminal/Command)>

Type:10h~1Fh (10h:Program NO.0, 11h:Program NO.1,

, 1F:Program NO.15)

Category:08h, Item:03h
Target1~4:00h

byte

Setting Data

0

Storage data amount: 0 to buffer size upper limit.

*Upper limit value is the max points can be set by LJ-

Navigator2.

1
2
3

<Storage data amount (OUT data (Edge))>

Type:10h~1Fh (10h:Program NO.0, 11h:Program NO.1,

, 1F:Program NO.15)

Category:08h, Item:04h
Target1~4:00h

byte

Setting Data

0

Storage data amount: 0 to buffer size upper limit.

*Upper limit value is the max points can be set by LJ-

Navigator2.

1
2
3
4

Threshold: The range that can be input in the OUT

minimum display unit. (0.01μm unit. Sined 32-bit

integer)

5
6
7
8

Hysteresis: 0 Hysteresis display range upper

limit for the minimum display unit. (0.01μm unit.

Sined 32-bit integer)

9

10
11

12

Target OUT: 0:OUT1, 1:OUT2,

2:OUT3

15:OUT16

13

Edgedirection: 0:Rising, 1:Falling

14

Reserved (fixed as 0)

15

Reserved (fixed as 0)

<Storage data amount (OUT data (level)>

Type:10h~1Fh (10h:Program NO.0, 11h:Program NO.1,

, 1F:Program NO.15)

Category:08h, Item:05h
Target1~4:00h

byte

Setting Data

0

Upper limit value: The range that can be input in

the OUT minimum display unit. (0.01μm unit.

Sined 32-bit integer)

1
2
3
4

Lower limit value: The range that can be input in

the OUT minimum display unit. (0.01μm unit.

Sined 32-bit integer)

5
6
7
8

Target OUT: 0:OUT1, 1:OUT2

15:OUT16

9

Reserved (fixed as 0)

10

Reserved (fixed as 0)

11

Reserved (fixed as 0)

<Program name>

Type:10h~1Fh (10h:Program NO.0, 11h:Program NO.1,

, 1F:Program NO.15)

Category:09h, Item:00h
Target1~4:00h

byte

Setting Data

0

Program name, byte0

1

Program name, byte1

2

Program name, byte2

~

~

23

Program name, byte23

* 24 characters max. 0 is not appended to the end.
* SHIFT-JIS

0:Peak, 1:bottom, 6:Average, 8:Maximum thickness,
9:Minimum thickness, 10:Average thickness, 11:Maximum
thickness position, 12:Mimimum thickness position,
13:Middle value, 14:P-P (Z) (only when profile compression
(time axis) is on),16:P-P (only for Height (simple 3D)

byte

Setting Data

0

Area Left: any value in measurement range

(0.01μm unit Sined 32-bit integer

example: 5mm=500000)

1
2
3
4

Area Right: any value in measurement range

(0.01μm unit Sined 32-bit integer

example: 5mm=500000)

5
6
7