KEYENCE LJ-V7000 Series User Manual

Page 85

Advertising
background image

85

LJ-V7000_COM_RM_E

5:Centerof circle

byte

Setting Data

0

Arc calculation areacount:

0: Area2 disabled, 1: Area2 enabled

1

Reserved (fixed as 0)

2

Reserved (fixed as 0)

3

Reserved (fixed as 0)

4

Arc calculation area Left:

any value in measurement range
(0.01μm unit Sined 32-bit integer

example: 5mm =500000)

5
6
7
8

Arc calculation area Right:

any value in measurement range
(0.01μm unit Sined 32-bit integer

example: 5mm =500000)

9

10
11
12

Area2 Left: any value in measurement range

(0.01μm unit Sined 32-bit integer

example: 5mm =500000)

13
14
15
16

Area2 Right: any value in measurement range

(0.01μm unit Sined 32-bit integer example: 5mm

=500000)

17
18
19

7:Edge (when profile compression (time axis) is off)

byte

Setting Data

0

Edge measuring area Left: any value in

measurement range

(0.01μm unit Sined 32-bit integer

example: 5mm=500000)

1
2
3
4

Edge measuring area Right: any value in

measurement range

(0.01μm unit Sined 32-bit integer

example: 5mm=500000)

5
6
7
8

Edge direction: 0:Rising, 1:Falling

9

Detect direction: 0:+direction, 1:–direction

10

Detect No: 1~10

11

Reserved (fixed as 0)

12

Edge level: Any value in measurement range

(0.01μm unit Sined 32-bit integer

example: 5mm=500000)

13
14
15

7:Edge (when profile compression (time axis) is on)

byte

Setting Data

0

Edge measuring area Left: any value in

measurement range

(0.01μm unit Sined 32-bit integer

example: 5mm=500000)

1
2
3
4

Edge measuring area Right: any value in

measurement range

(0.01μm unit Sined 32-bit integer

example: 5mm=500000)

5
6
7
8

Detection target: 0:Upper profile, 1:Lower profile

9

Edge direction: 0:Rising, 1:Falling

10

Detect direction: 0:+direction, 1:–direction

11

Detect No: 1~10

12

Edge level: any value in measurement range

(0.01μm unit Sined 32-bit integer

example: 5mm=500000)

13
14
15

15:P-P (X) (only for when profile compression (time axis) is
on)

byte

Setting Data

0

Edge measuring area Left: any value in

measurement range

(0.01μm unit Sined 32-bit integer

example: 5mm=500000)

1
2
3
4

Edge measuring area Right: any value in

measurement range

(0.01μm unit Sined 32-bit integer

example: 5mm=500000)

5
6
7
8

Edge direction: 0:Rising, 1:Falling

9

Detect direction: 0:+direction, 1:–direction

10

Detect No: 1~10

11

Reserved (fixed as 0)

12

Edge level: any value in measurement range

(0.01μm unit Sined 32-bit integer

example: 5mm=500000)

13
14
15

Advertising