[Unique parameters of measurement mode from
4byte]

Measurement mode 0:OFF

byte

Setting Data

4

Reserved (fixed as 0)

~

91

1:Height (when profile compression (time axis) is off)

byte

Setting Data

4

Measurement target head: 0: HeadA, 1: HeadB, 2:

Combined profile (only when profile combine

(wide) is on)

5

Measurement target selection: 0: Peak, 1: Bottom,

2: Knee, 3: Intsect (lines), 4: Intsect (lin-arc), 5:

Center of circle, 6: Average

6

Position correction selection: 0: No position

correction, 1: Position correction1, 2: Position

correction2

7

Reserved (fixed as 0)

8

Unique parameters are assigned to each

measurement target. For details on the unique

parameters, see "Measurement area details"

(page 83).

~

N

N+1

Reserved (fixed as 0)

~

91

2:Step

byte

Setting Data

4

Measurement target head: 0: HeadA, 1: HeadB, 2:

Combined profile (only when profile combine

(wide) is on)

5

Measurement target selection: 0: Peak, 1: Bottom,

2: Knee, 3: Intsect (lines), 4: Intsect (lin-arc), 5:

Center of circle, 6: Average

6

Position correction selection: 0: No position

correction, 1: Position correction1, 2: Position

correction2

7

Reserved (fixed as 0)

8

Unique parameters are assigned to each

measurement target. For details on the unique

parameters, see "Measurement area details"

(page 83).

~

M

M+1

Reserved (fixed as 0)

~

47

48

Reference target head: 0: HeadA, 1: HeadB, 2:

Combined profile (only when profile combine

(wide) is on)

49

Referene target selection: 0:Peak, 1:Bottom,

2:Knee, 3:Intsect (lines), 4:Contact (lin-arc), 5:

Center of circle, 6: Average

50

Position correction selection: 0: No position

correction, 1: Position correction1, 2: Position

correction2

51

Reserved (fixed as 0)

52

Unique parameters are assigned to each

Reference target. For details on the unique

parameters, see "Measurement area details"

(page 83).

~

N

N+1

Reserved (fixed as 0)

~

91

3:Position (when profile compression (time axis) is off)

byte

Setting Data

4

Measurement target head: 0: HeadA, 1: HeadB, 2:

Combined profile (only when profile combine (wide) is on)

5

Measurement target selection: 0:Peak, 1:Bottom,

2:Knee, 3:Intsect (lines), 4:Contact (lin-arc),

5:Center of circle, 7:Edge

6

Position correction selection: 0: No position correction,

1: Position correction1, 2: Position correction2

7

Reserved (fixed as 0)

8

Unique parameters are assigned to each

measurement target. For details on the unique

parameters, see "Measurement area details"

(page 83).

~
N

N+1

Reserved (fixed as 0)

~

91

4:Center position, 5:Width

byte

Setting Data

4

Measurement target head: 0: HeadA, 1: HeadB, 2:

Combined profile (only when profile combine (wide) is on)

5

Measurement target selection: 0:Peak, 1:Bottom,

2:Knee, 3:Intsect (lines), 4:Contact (lin-arc),

5:Center of circle 7:Edge

6

Position correction selection: 0: No position correction,

1: Position correction1, 2: Position correction2

7

Reserved (fixed as 0)

8

Unique parameters are assigned to each

measurement target. For details on the unique

parameters, see "Measurement area details"

(page 83).

~

M

M+1

Reserved (fixed as 0)

~

47

48

Reference target selection: 0: HeadA, 1: HeadB, 2:

Combined profile (only when profile combine (wide) is on)

49

Reference target selection: 0:Peak, 1:Bottom,

2:Knee, 3:Intsect (lines), 4:Contact (lin-arc),

5:Center of circle, 7:Edge

50

Position correction selection: 0: No position correction,

1: Position correction1, 2: Position correction2

51

Reserved (fixed as 0)

52

Unique parameters are assigned to each

Reference target. For details on the unique

parameters, see "Measurement area details"

(page 83).

~
N

N+1

Reserved (fixed as 0)

~

91

6:Thickness

byte

Setting Data

4

Measurement target selection: 8:Max thickness,

9:Min thickness, 10:Ave thickness, 11:Max

thickness position, 12:Min thickness position

5

Position correction selection (HeadA): 0: No position

correction, 1: Position correction1, 2: Position correction2

6

Position correction selection (HeadB): 0: No position

correction, 1: Position correction1, 2: Position correction2

7

Reserved (fixed as 0)

8

Unique parameters are assigned to each

measurement target. For details on the unique

parameters, see "Measurement area details"

(page 83).

~

N

N+1

Reserved (fixed as 0)

~

91