KEYENCE LJ-V7000 Series User Manual

Page 75

Advertising
background image

75

LJ-V7000_COM_RM_E

• OUT settings

<Profile Mask area settings individual>

Type:10h~1Fh (10h:Program NO.0, 11h:Program NO.1,

, 1F:Program NO.15)

Category:05h, Item:02h
Target1:00h (headA/wide), 01h (headB)
Target2:00h~04h (Profile mask area1~5) Target3~4:00h

byte

Setting Data

0

Area selection:0:Disabled, 1:Rectangle, 2:Triangle

1

position correction selection: 0: No position

correction, 1:Position correction1, 2:Position

correction2

2

Reserved (fixed as 0)

3

Reserved (fixed as 0)

4

Xcoordinate1: any value in measurement range

(0.01μm unit Sined 32-bit integer

example: 5mm=500000)

5
6
7
8

Zcoordinate1: any value in measurement range

(0.01μm unit Sined 32-bit integer

example: 5mm=500000)

9

10
11
12

Xcoordinate2: any value in measurement range

(0.01μm unit Sined 32-bit integer

example: 5mm=500000)

13
14
15
16

Zcoordinate2: any value in measurement range

(0.01μm unit Sined 32-bit integer

example: 5mm=500000)

17
18
19
20

Xcoordinate3: any value in measurement range

(0.01μm unit Sined 32-bit integer

example: 5mm=500000)

21
22
23
24

Zcoordinate3: any value in measurement range

(0.01μm unit Sined 32-bit integer

example: 5mm=500000)

25
26
27

* When Rectangle is selected, upper left (Xcoordinate1,

Zcoordinate1) and bottom right (Xcoordinate2,
Zcoordinate2) should be set. Xcoordinate3, Zcoordinate3 is
no effect.

* When Triangle is selected, (Xcoordinate1, Zcoordinate1),

(Xcoordinate2, Zcoordinate2) and (Xcoordinate3,
Zcoordinate3) are used.

<OUT name>

Type:10h~1Fh (10h:Program NO.0, 11h:Program NO.1,

, 1F:Program NO.15)

Category:06h, Item:01h
Target1:00h~0Fh (OUT1~16) Target2~4:00h

Byte

Setting Data

0

OUT name, byte1

1

OUT name, byte2

2

OUT name, byte3

~

~

19

OUT name, byte20

*20 Characters max. 0 is not appended to the end.
*SHIFT-JIS

<Minimum display unit>

Type:10h~1Fh (10h:Program NO.0, 11h:Program NO.1,

, 1F:Program NO.15)

Category:06h, Item:02h
Target1:00h~0Fh(OUT1~16) Target2~4:00h

byte

Setting Data

0

Minimum display unit

1

Reserved (fixed as 0)

2

Reserved (fixed as 0)

3

Reserved (fixed as 0)

* The unit changes according to the measurement mode

assigned to the OUT.

Length system, 0:1mm, 1:0.1mm, 2:0.01mm, 3:0.001mm,
4:1μm, 5:0.1μm
Area system, 0:1mm

2

, 1:0.1mm

2

, 2:0.01mm

2

, 3:0.001mm

2

,

4:0.0001mm

2

, 5:0.00001mm

2

Angle system, 0:1deg, 1:0.1deg, 2:0.01deg

<Measurement mode>

Type:10h~1Fh (10h:Program NO.0, 11h:Program NO.1,

, 1F:Program NO.15)

Category:06h, Item:03h
Target1:00h~0Fh (OUT1~16) Target2~4:00h

byte

Setting Data

0

Minimum display unit (from

)

1

Measurement mode (from

)

2

Reserved (fixed as 0)

3

Reserved (fixed as 0)

4

From 4byte, unique parameters are assigned to

each measurement mode. For details on the

unique parameters, see [Unique parameters of

measurement mode from 4byte]

~

N

Length system

0:1mm, 1:0.1mm, 2:0.01mm,

3:0.001mm, 4:1μm, 5:0.1μm
Area system

0:1mm

2

, 1:0.1mm

2

, 2:0.01mm

2

,

3:0.001mm

2

, 4:0.0001mm

2

, 5:0.00001mm

2

Angle system 0:1deg, 1:0.1deg, 2:0.01deg

0:OFF, 1:Height, 2:Step, 3:Position, 4:Center position,
5:Width, 6:Thickness, 7:Angle, 8:R measurement, 9:Area,
10:Master comparison (Z), 11:Distance(point-point),
12:Distance(point-line), 13:Heigh (Profile compression
(time axis) on), 14:Position (Profile compression (time axis)
on), 15:Deflection width (Profile compression (time axis)
on), 16:Height (simple 3D), 17:Step (simple 3D),
18:Position (simple 3D), 19:Calculation

Advertising