1 bl-180 connections, 1 wire colors and signal types, 2 power supply wiring – KEYENCE BL-180 User Manual

Page 10: 3 wiring i/o

Advertising
background image

1.1 BL-180 Connections

2

1.1

BL-180 Connections

1.1.1

Wire colors and signal types

The following wires extend from the BL-180. Solder the required wires to a con-
nector to connect the BL-180 to a computer/controller.

1.1.2

Power supply wiring

• Be sure to match the polarities of the power supply when soldering the con-

nections. Reversing the polarities will damage the unit.

• Make sure that the power supply provides a stable 5 VDC ± 5%. If the power

supply does not function in the above range, it can damage the unit.

• Do not use a power cable longer that 2 meters. A long power cable can cause

a voltage drop, preventing the BL-180 from starting properly.

• If the power supply is UL rated, it must provide Class 2 output.

1.1.3

Wiring I/O

Trigger (TIM) input

The trigger input is used to signal the BL-180 to start reading (Start laser emis-
sion).
The trigger input is a non-voltage input (TTL input is also available).

Wire Color

Symbol

Description

Signal Direction

Shield

Shield

Connect to ground (SG)

——

Purple

RS-
232C

SD (TXD) Send data

Output

Brown

RD (RXD) Receive data

Input

Pink

RS (RTS) Request to send (always on)

Output

Blue

CS (CTS) Request to receive

Input

Black

GND (SG)

Ground (common ground for
respective signals)

——

Yellow

TIM

Trigger input

Input

White

OK

OK output

Output

Gray

NG

NG output

Output

Red

+ 5V

+ 5V power supply input

Input

5VDC

BL-180

+5V

Red

GND

Black

+

CAUTION

GND

TIM

Yellow

Black

5VDC

10K

4.7
K

BL-180

Internal circuit

With or
without relay

Advertising