3 setting the bl-u1 dip switches, Power supply bl-u1, Caution – KEYENCE BL-180 User Manual

Page 90

Advertising
background image

Power Supply BL-U1

84

Prepare the BL series for connection to the BL-U1 by soldering a D-sub 9-pin con-
nector to the BL series cable. Then connect the cable to READER port of the BL-
U1.

Prepare the D-sub 9-pin connector and its connector case separately.

Take special care when soldering pin 5 (GND) and pin 9 (+5VDC). A wrong con-
nection will damage the unit.

Do not use a power cable over 2 meters long. A long power cable can cause a
drop in voltage, preventing the BL series from starting up properly.

1.3

Setting the BL-U1 DIP switches

According to the selected interface and timing input, change the DIP switch set-
tings.

Shield

Yellow

Blown

Purple

White

Black

Gray

BL series

1

2

-

3

5

6

4

7

Connector case

TIM

RD

SD

OK

GND

NG

RS

Pink

Blue

Red

8

9

CS

+5VDC

D-sub 9-pin (female)

#4-40 screw

READER port

BL-U1

Use a metallic connector case for the D-sub 9-pin connector and connect the
shielded line to the connector case. This allows connection to the earth ground of
the AC power cable.

CAUTION

OFF

ON

1

2

3

4

5

6

The switch settings at left are the
factory default settings.

DIP switch setting

1

2

3

4

5

6

Interface select

RS-232C

ON

OFF

OFF

RS-422A

OFF

ON

OFF

RS-485 multidrop

OFF

OFF

ON

RS-422A terminator
(terminal resistance 100

)

OFF

OFF

ON

ON

RS-485 terminator
(terminal resistance 100

)

OFF

OFF

ON

ON

READER port CS control
method select

Reflect ON/OFF of CS
at RS-232C port

OFF

Always ON

ON

Advertising