4 function and wiring on the i/o terminal block, Power supply bl-u1, Wiring the trigger input – KEYENCE BL-180 User Manual
Page 91
Power Supply BL-U1
85
1.4
Function and wiring on the I/O terminal block
The above list starts from the left edge terminal on the terminal block and sequen-
tially goes to the right.
• M 3.0 screws are used on the terminal block.
• Use the clips shown below when wiring.
Wiring the trigger
input
The trigger input signals the BL series to start reading.
Turn on the trigger input by supplying 8.5 to 30 VDC between the trigger input ter-
minals.
You can use the “power supply for sensors” terminals of the BL-U1 as the input
power supply.
Symbol
Description
Signal Direction
TIM
Timing input
Input
Input
+12V OUT
Power supply for sensors (+) (+12 VDC,
300 mA)
Output
Power supply for sensors (-) (0 V)
Output
COM
Common for OK/NG output
——
OK
OK output
Output
NG
NG output
Output
SDA
Send RS-422A data (+)/RS-485 (+)
Output/Input and output
SDB
Send RS-422A data (-)/RS-485 (-)
Output/Input and output
SG
Signal ground
——
RDA
Receive RS-422A data (+)
Input
RDB
Receive RS-422A data (-)
Input
TIM
+12V OUT– COM
OK
NG
SDA
SDB
SG
RDA
RDB
Trigger
input
Power supply
for sensors
(12 VDC, 300 mA)
OK/NG output
RS-422A/RS-485
6.0 mm or
less
Round-shape
6.0 mm or
less
Fork-shape
TIM
+12V OUT–
8.5~30VDC
+
+
Contact or
non-contact