3.3 Setup Software Operating Procedure

34

4. When you select Inspect check-digit, the following information appears:

The above setting sends the data together with the check-digit.
Modulus 43 is used to calculate the check-digit.

5. When you select Max code length output, the following information appears:

Set Effective and Starting from 1 to 32.

6. When you select Specify label orientation, the following information appears:

7. Click the

Return

button or press

[ENTER]

to return to the main setting

screen.

Max. code length output --> Page 23
Specify label orientation --> Page 16

3.3.4

ITF setup

1. Setup is the same as the CODE39 setup. However, ITF does not have Start/

stop character.

2. Set code length using a even value within the range from 2 to 32.

3. Modulus 10/Weight3 is used to calculate the check-digit.

When reading the standard distribution code (bar code on the corrugated fiber-
board box), set the code length to 14 or 16 and select the Inspect check-digit.

3.3.5

Setup for Industrial 2 of 5 and COOP 2 of 5

1. Setup is the same as the CODE39 setup. However, a start/stop character

and Inspect check-digit are not provided.

2. Set code length within the range from 1 to 32.