2. during use – A&D Weighing GF-32K User Manual

Page 10

Advertising
background image

8

The meaning of IP-65 is "No ingress of dust. Projected against water jets".

If a powerful water jet is used or the balance is immersed in water, it may cause a damage
that is due to ingress of water.

Confirm that "the plug is inserted firmly into the jack" and "the terminal is covered using the

waterproof cover or the waterproof RS-232C cable (GX-07K)", when using the balance.

Use the waterproof option cable GX-07K, when the RS-232C interface is used with IP-65.

AX-KO1710-200, a standard RS-232C cable, is not waterproof or dustproof.

Confirm that the weighing pan does not touch to rim.

Do not install the balance where flammable or corrosive gas is present.


3-2. During Use

Discharge static electricity from the weighing material. When

weighing sample (plastics, insulator, etc.) could have a static
charge, the weighing value is influenced. Ground the balance, and

Eliminate the static electricity by AD-1683 as an accessory.

Or try to keep the ambient humidity above 45%RH at the room.

Or use the metal shield case.

Or wipe a charged material (plastic sample etc.) with the wet cloth.

This balance uses a strong magnet as part of the balance

assembly, so please use caution when weighing magnetic
materials such as iron. If there is a problem, use the underhook on
the bottom of the balance to suspend the material away from the
influence of the magnet.

Eliminate any temperature difference between the sample and the

environment. When a sample is warmer (cooler) than the ambient
temperature, the sample will be lighter (heavier) than the true
weight. This error is due to a rising (falling) draft around the sample.

Make each weighing gently and quickly to avoid errors due to

changes in the environmental conditions.

Do not drop things upon the weighing pan, or place a sample on

the pan that is beyond the balance weighing capacity. Place the
sample in the center of the weighing pan.

Do not use a sharp instrument such as a pencil to press the keys.

Use your finger only.

Press the

RE-ZERO

key before each weighing to prevent

possible errors.

Calibrate the balance periodically so as to eliminate possible errors.
Take into consideration the affect of air buoyancy on a sample

when more accuracy is required.

It is possible to check the reference card for principle operation.

Material case

Charged

sample

Magnetic

material

40°C

20°C

Draft

Weighing Pan

Grounding

Shock

Advertising
This manual is related to the following products: