8-3. adding the comparison results, 8-4. main display comparison function – A&D Weighing GF-32K User Manual

Page 52

Advertising
background image

50

10-8-3. Adding the Comparison Results

By setting the "Comparison results (

Cp-r

)" of the function table to "

1

", the comparison results

can be added to the data output using the RS-232C serial interface. Use A&D standard
format (

type 0

). The AD-8121B printer can not be used.

The comparison results are added after the header in A&D standard format as below.











10-8-4. Main Display Comparison Function

The main display comparison function displays the comparison results in a magnified way,
on the main portion of the display in place of the weight value.

Selecting a unit

Step 1 Press

the

MODE

key to select a unit to be used for comparison.

Note While the main display comparison function is in use, unit selection using

the

MODE

key is not available.

Setting the function table

Step 2 Press and hold the

SAMPLE

key until

ba5fnc

of the function table is displayed,

then release the key.

Step 3 Press

the

SAMPLE

key several times to display

Cp fnc

.

Step 4 Press

the

PRINT

key.

Step 5 Press

the

SAMPLE

key several times to display

Cp-b 0 .

Step 6 Press

the

RE-ZERO

key to display

Cp-b 1

.

Note To disable the main display comparison function, set the "Main display

comparison (Cp-b)" parameter to "0".

Step 7 Press

the

PRINT

key to store the setting.

Step 8 Press

the

CAL

key to return to the weighing mode.

Setting the comparator values

Setting the comparator values as described in the previous section.

This example uses

Cp 3

(Continuous comparison, excluding "near zero").

Using the main display comparison function

Step 1 Press

the

RE-ZERO

key to set the display to zero.

Step 2

Place a sample on the pan. The balance performs a comparison using the specified
comparison values and displays the comparison results, HI, OK or LO.

Header Data

Unit

Terminator

Comparison

result

When the comparison result is HI
When the comparison result is OK
When the comparison result is LO
Not applicable

Advertising
This manual is related to the following products: