8-3. One-Touch Calibration For The GX-K Series

Calibration using the internal mass for The GX-K series

This function calibrates the balance using the internal mass. The only operation required is
to press the

CAL

key.

Caution

GF-K series can not use this calibration mode.

Step 1 Connect the AC adapter and warm up the balance for at least 30 minutes with

nothing on the weighing pan.

Step 2

Press the

CAL

key.

Step 3

The balance displays

Calin

and performs calibration using the internal mass. Do

not allow vibration or drafts to affect the balance.

Step 4

The balance displays

end

after calibration.

If the

"GLP output (

info

)" parameter of

the function table is set to "

1

" or "

2

", the balance displays

glp

and outputs the

"calibration report" using the RS-232C interface or stores the data in memory. Refer
to "11-2. GLP Report" and "Data memory (

data

)" of the function table for details.

Step 5 The balance will automatically return to the weighing mode after calibration.

About the internal mass

The value of the internal mass may change due to aging, corrosion or other damage caused
by the operating environment. Check the internal mass periodically. Correct the internal
mass value as necessary. Refer to "8-6. Correcting the internal mass value".

To maintain the weighing accuracy, perform the calibration using an external weight

periodically, as described below.