A&D Weighing GF-32K User Manual

Page 32

Advertising
background image

30

CAL

key..........To cancel the correction and display the next

menu item of the function table.

Step 10 Press the

CAL

key to return the weighing mode.




Step 11 Press the

CAL

key to calibrate the balance using the internal mass.

Step 12 Place the external weight on the pan and confirm that the correction has been

performed properly. In this example, confirm that the value displayed is within the
range that is described at "Accuracy after calibration using the internal mass" of "20.
Specification ". If the value is incorrect, repeat the correction.