14. Programmable Unit

This is a programmable unit conversion function. It multiplies the weighing data in kilograms by an
arbitrary coefficient set in "Programmable unit (

mlt

)" of the function table and displays the result.

The coefficient must be within the range between the minimum and maximum shown below. If the
coefficient set is beyond the range, an error is displayed and the balance returns to the coefficient
setting mode, prompting to enter an appropriate value. A coefficient of 1 was set at the factory.

Model Minimum

coefficient Maximum coefficient

GX-8K

GX-10K

GF-8K

GF-10K

100000

GX-8K2 GX-12K GX-20K GX-30K
GF-8K2 GF-12K GF-20K GF-30K

10000

GX-32K

GF-32K

0.000001

1000

Operation

Step 1 Press and hold the

SAMPLE

key until

ba5fnc

of the function table is displayed,

then release the key.

Step 2 Press

the

SAMPLE

key several times to display

mlt

.

Step 3 Press

the

PRINT

key to enter the mode to confirm or set the coefficient.


Confirming the coefficient
Step 4 The current coefficient is displayed with the first digit blinking.

When it is not to be changed, press the

CAL

key and go to step 6.

When it is to be changed, press the

RE-ZERO

key and go to step 5.


Setting the coefficient
Step 5 Set the coefficient using the following keys.

SAMPLE

key ....To select a digit to change the value.

The selected digit blinks.

RE-ZERO

key...To change the value.

MODE

key .......To change the decimal point position.

Each time the switch is pressed, the
decimal point position changes as follows:



PRINT

key........To store the new coefficient, display

end

and go to step 6.

CAL

key...........To cancel the new coefficient and go to step 6.


Quitting the operation
Step 6 The balance displays

Unit

. Press the

CAL

key to exit the programmable-unit

function and return to the weighing mode.


Using the function
Press

the

MODE

key to select the programmable-unit (no display on the unit section).

Perform weighing as described in "6-1. Basic Operation (Gram Mode)". After weighing, the
balance displays the result (weighing data in kilograms

x

coefficient).

Store