A&D Weighing GF-32K User Manual

Page 67

Advertising
background image

65

Notes

Using the "CN:mm" command, the comparator limit values can be recalled.

The upper limit value recalled can be output using the "?HI" command.

The lower limit value recalled can be output using the "?LO" command.

The upper limit value can be changed using the "HI:" command.

The lower limit value can be changed using the "LO:" command.

"mm" indicates a two-digit numerical value 01 to 20, which corresponds to C01 - C20.


Digital input mode

In the digital input mode, the upper and lower limit values are
entered digitally using the keys. Use the following keys to store
upper and lower limit values in digital input mode.

SAMPLE

key ... To select the digit to be changed.

RE-ZERO

key.. To change the value of the selected digit.

MODE

key ...... To switch the polarity.

PRINT

key....... To store the upper and lower limit values.

Go to step 3 of the previous page.

CAL

key.......... To return to the upper and lower limit

values confirmation mode

Go to step 3 of the previous page.

MODE

key

Press and hold to go to the weighing input mode.


Weighing input mode

In the weighing input mode, a sample is placed on the pan to store the upper and lower limit
values.


Notes

Pressing

the

CAL

key will interrupt the operation and the

balance will return to the upper and lower limit values
confirmation mode (step 3 in "Storing the upper and lower
limit values").

To go to the digital input mode, press and hold the

MODE

key.

Step 1 The first display in the weighing input mode depends on the

comparator number selected in step 4 in "Storing the upper and
lower limit values".

For example, when "

C03 H1

" is selected in step 4, the display is

the current weight value and the comparator number with

HI

blinking.

Step 2 Place a container on the weighing pan, if necessary.

Press

the

RE-ZERO

key to set the display to zero.

Step 3 Place a sample corresponding to the upper limit value,

on the pan or in the container.

Step 4 Press the

PRINT

key to store the upper limit value.

Value Digit

Polarity

Store

C03

HI

Upper limit value

Comparator

number

Upper

limit

Store

Place a sample

corresponding to the

upper limit value

Advertising
This manual is related to the following products: