1-2. measuring the density – A&D Weighing GF-32K User Manual

Page 78

Advertising
background image

76

15-1-2. Measuring the density

Note If the liquid temperature or the type of liquid is changed during measurement,

input again the value of the liquid density as necessary. Refer to "C. Entering
the density of a liquid".

Entering the density mode

Step 1 Press

the

MODE

key as necessary to select the density mode.

When the density mode begins, "

d-a

" (weight in air), "g (gram)" for the unit and

the processing indicator ( ) are turned on.

Step 2 Selecting a parameter to set.

Press

the

SAMPLE

key to select a parameter to set. Each pressing of the key

switches the parameter.

Press

the

RE-ZERO

key to enter the inputting mode of the parameter selected.



















Setting the parameter

Step 3 Setting the parameter.

(A) Entering the weight of a sample in air

1 Confirm that nothing is placed on the upper pan (in air).

Press the

RE-ZERO

key to set the display to zero.

2 Place a sample on the upper pan. When the value

displayed on the balance becomes stable, press the

PRINT

key. The sample weight in air is confirmed and

the balance will return to the density mode (Go to "Step
2 Selecting a parameter to set").

3 Remove the sample from the upper pan.


Note If an unstable value, a negative value or

e

(

error

)

is displayed, the

PRINT

key is disabled.

The

SAMPLE

key can not be used to change the

minimum weighing value.

Press

Put sample

New parameter

Press

To other

weighing units

Weight in air

Weight in

liquid

Liquid

Density

Calculated

result

Density mode

Input mode

(A) Entering the weight

of a sample in air

(B) Entering the weight

of a sample in liquid

(C) Entering the density

of a liquid

Pressing the SAMPLE

key to display the result

Advertising